KRSS (formerly Kinesis-RSS) has helped companies across the globe optimise, move, expand and replace LC-MS/MS & UHPLC-MS systems. Please read our latest techtips below to see how the newest approaches and innovations can help you transform your cost and service challenges into real business advantages.

How to properly install sample loops

Read More
Stainless SteelStainless steel sample loops are supplied with fittings that are unswaged onto the tube. It is important that the loop be completely bottomed in the injector port before the ferrule is ...

Avoiding Pressure Transients

Read More
Air in the sample loop can cause instantaneous system pressure drop that eventually returns to a normal level. Air causes the pressure to drop when the injector moves from the LOAD to the INJECT posit...

How to properly install fittings

Read More
RheFlex® Stainless Steel (and titanium) A good square-cut, burr-free end on tubing is important for use in your system. The square-cut makes a flat surface for the tube to make contact against the bot...

Sample Loop Loading

Read More
Partial-FillingUse the partial-filling method if you need to conserve sample or if you want to vary sample volume frequently. In partial-filling, the syringe sets the volume injected onto the column. ...

How to find and fix common leaks

Read More
Leaks cause valuable sample loss. Nobody wants that! The key to the valve holding pressure is the integrity of the sealing surfaces. If there is a scratch on the sealing surface or the needle seal in ...

How to use proper syringe needles

Read More
With front-loading injectors, it is important to use the correct needle when loading the sample loop. An incorrect needle will damage the valve and can cause poor reproducibility. When the needle is t...